Conductivity |
|
4 measuring ranges – Quadropole electrode
Temperature compensation is user-selectable depending on the brand of dialysis machine
Range: |
0 to 30.00 mS/cm |
Cell constant: |
typical 0,54 1/cm |
Accuracy: |
0 to 199 uS/cm ± 0.6 uS/cm |
|
200 to 1999 uS/cm ± 6 uS/cm |
|
2 to 11.99 mS/cm ± 0.06 mS/cm |
|
12 to 15.99 mS/cm ± 0.03 mS/cm |
|
16 to 30 mS/cm ± 0.06 mS/cm |
With option C, the conductivity/temperature probe may be disconnected.
See picture above.
This option is available for all meters.
|
Temperature |
|
Sensor integrated in the conductivity electrode
Range: |
0…100 °C |
Resolution: |
0.01 °C |
Accuracy: |
± 0.1 °C |
|
Pressure |
|
Reading selectable as mmHg, kPa, Bar or PSI
Version HDM97BQ and BO |
Range: |
|
-700 to + 1900 mmHg, 2 Bar |
Resolution: |
|
0.1 mmHg |
Accuracy: |
|
0 … 300 mmHg ± 1 mmHg
otherwise ± 2 mmHg |
|
|
in temperature range 20 to 23 °C
Ceramic sensor fluid resistant |
Option H with higher accuracy |
Range: |
|
-700 to + 1900 mmHg, 2 Bar |
Resolution: |
|
0.01 mmHg |
Accuracy: |
|
0 … 300 mmHg ± 0.5 mmHg,
1000 … 1900 mmHg ± 2 mmHg
otherwise ± 1 mmHg |
|
|
in temperature range 20 to 23 °C
Sensor not fluid resistent |
|
Flow Only Version BQ |
|
Range: |
100 … 2000 ml/min |
Resolution: |
1 ml/min |
Accuracy: |
± 2 % from reading |
|
Power Supply |
|
Li-Ion-Polymer battery, rechargeable with 5 V DC, 500 mA
External charger:
Input Voltage 100 … 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates up to 200 hours with fully charged battery |
Dimensions and Weight |
|
165 x 82 x 33 mm – 6.5″ x 3.2″ x 1.3”
0.4 Kg – 0.88 lbs including probe |